GMI Electroacoustic Ear (EARS)

£425.00 ex. VAT

EARS (Electroacoustic Auto Response System) has been designed to provide a quick and convenient way of checking the performance of an audiometer, and features a unique response button testing facility.

Electroacoustic ears provides a quick and convenient way of checking the performance of an audiometer and the response button prior to using it with subjects. This electroacoustic ear provides accurate and consistent responses to tones and gives thresholds over the frequency range 500Hz – 8000Hz.

What can EARS offer you?

EARS provides you with the ability to quickly confirm that the sound output of your audiometer, as delivered to the earphones or through booth connections, is as expected. It also has a unique response button testing facility.

How does it do these things?

The earphones are fitted onto the colour coded mouldings, as you would to a person under test, and a cable from Ears is plugged into your audiometer response socket in place of your response button. Then run a full automatic or manual test and compare the thresholds achieved with the target values or previous results recorded.

Your response button can be plugged into the socket marked Response Unit Test on the EARS unit and correct operation confirmed by pressing and releasing the button.

Why check your audiometer with an electroacoustic ear?

Checking performance prior to use is recommended by the Health & Safety Executive in their publication ‘Controlling Noise at Work’, in order to pick up degradation in booth sockets, cables, audiometer earphones and audiometer performance. EARS is the ideal choice for performing this check as it consistently provides accurate thresholds.

Product Features & Specifications

  • Frequency Range: 500Hz – 8000Hz
  • Unique response button testing facility
  • Fully automatic
  • Accurate & Consistent
  • USB-powered, saving on costly batteries
  • Size: 120 x 120 x 94mm
  • Weight: 450g

Thresholds

  • 500Hz – 85dB
  • 1000Hz – 80dB
  • 2000Hz – 80dB
  • 3000Hz – 80dB
  • 4000Hz – 80dB
  • 6000Hz – 80dB
  • 8000Hz – 75dB

The Electroacoustic ear is compatible with the Asra Audiometer and the Osicus Audiometer

You can view the manufacturer information on the electroacoustic ears here.

This product is supplied with a 12 month calibration certificate.